YSAV604 Anesthesia Machine, with its ergonomic design, integrates a concise and smooth structure and high mobility. The microcomputer-controlled Ventilator adopts exported sensors and other components. Its scientific and proper design as well as rigorous production process demonstrate the inner quality: mechanical ventilation is made secure and visible by the folded respiratory bellows, and the security of air supply is guaranteed by pneumatic transmission and measurement system. With multiple modern technologies applied, the Anesthesia Machine is both outstanding in performance and versatile in functions.
Main body
High-strength engineering plastic rack, light,beautiful and corrosion resistant
Scope of application
adult, child
Gas source
O2: 0.27~0.55MPa N2O: 0.27~0.55MPa
Flow meter
O2: 0.1~1.0L/min 1.1~10L/min
N20: 0.1~1.0L/min 1.1~10L/min
O2, N2O linkage and N2O Stopper
When using nitrous oxide, oxygen concentration>25%;
When the oxygen pressure<0.2Kpa, the flow of nitrous oxide would be cut off.
Flow rate of rapid oxygen supply
25~75L/min
Low oxygen pressure alarm
There will be sound alarm when the oxygen
pressure < 0.2MPa
Vaporizer
it has the function of automatic compensation
based on pressure, temperature, and flow rate.
The regulation range of evaporator
concentration is 0~5 vol%.
Among Halothane,Enflurane,Isoflurane and
Sevoflurane,two can be chosen for application
as required by the customer. And vaporizers that
are imported with original packaging are also
available.
Respiratory circuit
working mode: all-close, semi-close, semi-open
APL: 0.5~7 kPa
Respiratory Bellows
bellows for adults, bellows for children
tidal volume range:0~1,500 ml
Ventilator
Display mode
High-definition 5.7” LCD screen display
Ventilation mode
IPPV, SIPPV, SIMV, PEEP, MANUAL, SIGH
Ventilation parameters
Tidal volume
50~1500 ml
Rate
2~99 bpm
SIMV rate
2~20 bpm
I:E
2:1~1:8
Inspiratory trigger pressure
-1.0~2.0 kPa
PEEP
0~2.0 kPa
Pressure range
1.0~6.0 kPa
SIGH
1.5 times the inspiratory time
Parameters for ventilation monitoring
tidal volume, ventilation volume, IPPV rate,SIMV rate, total respiratory rate,
I/E, peak pressure of airway, pressure - time waveform, flow rate - time
waveform, PEEP, inspiratory trigger pressure
Monitoring of oxygen concentration
(21%~100%)
Security Alarm System
Oxygen Concentration alarm
upper limit setting range (50%~100%)
low limit setting range (15%~50%)
Airway pressure alarm
upper limit setting range 1.0~6.0 kPa
low limit setting range 0.4~2.0 kPa
Per-minute ven