1. Microprocessor based servo controlled temperature system.
2.The microprocessor-controlled unit allows for three settings: pre-warm, manual or automatic.
3. Alarm functions for safety includes: Power failure, temperature deviation, Temperature sensor failure , over temperature.
4. APGAR timer to record therapy time
5. Convenient X-ray cassette under infant bed
5. Heater head can be adjusted + 90º horizontally
6. Tilt angle of infant bed is adjustable
7. Observe light angle is adjustable
8. Transparent protector folded
9. Adopts anti-explosion micro crystal quartz infrared radiation tube for its radiation source, the heating is quick and life-span is long
10. Alarms are audible and visual
11. A second thermal cut-out function for more safety
12. Skin temperature sensor failure protect function avoid over temperature
13. Temperature deviation can be amended in front panel directly
14. Low pressure aspirator, oxygen supply system and LED phototherapy unit
Technical data:
1. Power supply : 110/220V 60/50Hz
2. Power input: ≤1200VA
3. Operating condition
An ambient Temperature:18°C-30°C
An ambient Relative Humidity: 30%--75%RH
Atmospheric Pressure: 700~1060 hPa
An ambient Air Velocity <0.3m/s
4. Temperature control: 32°C~38°C
5. Temperature control accuracy: ≤0.5°C
6. Mattress temperature uniformity : <2°C
7. Skin temperature sensor accuracy: + 0.3°C
8. Warm-up time: ≤30min
9. Infant Bed tilt angle:+ 10°
10. APGAR timer range: ≤24h
11. Heater head horizontal: ±90°
12. Mattress size: 67CM X 54CM
13. Distance from heater to mattress:90CM
14. Height adjustable range:0-20CM
15. LED phototherapy light: 2pcs X 8W
Transport and storage
Environment Temperature:-40°C~ +55°C
Environment Relative Humidity: ≤95%
Atmospheric Pressure: 500~1060hPa
Package:
Each unit is packed one case
Size:135CMX 85CM X 98CM Gross weight: ≤170KG