We use cookies to improve your online experience. By continuing browsing this website, we assume you agree our use of cookies.

Patient Monitors

Categories

YSPM-F15M Multi-parameter Modular Patient Monitor (15 inches)

Model: YSPM-F15M
  • ModelYSPM-F15M
  • BrandYSENMED
  • PriceGet Price
  • MOQ1
  • Ship FromFrom Guangzhou, China
  • ShippingAny shipping terms convenient for customers
  • Share:
    Description Get Price Review(0)

    Technical Data Sheet - Release V2.2

     

    The  F15 patient monitor is designed to match the pace and unique needs of adult, pediatric, and neonatal intensive care; anesthesia and peri-operative care; OR and cardiac care environments.

    An optional USB Mouse/Keyboard control eases navigation and data entry.

    An optional HDMI output eases observation from a long distance.

    Features

    Core

    Newly advanced A9 main board with Linux OS, calculating speed is 4 times faster than traditional products

    Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups NIBP review and 2000 alarm events review

     

    Body

    15 inch high-brightness TFT LED

    Support display 9~13 waveforms

    Support 7 channel ECG waveform display simultaneously  

    Optional touch screen

    Optional HDMI output

     

    Printer

    Built-in high-speed 50mm thermal printer

     (Brand: SEIKO, Japan)

     

    Central System

    Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station.

    Optional Support HL7 (Health Level Seven)

    Alarm

    Three-level acousto-optic alarm

    Sensor-off alarm

    Paper out alarm

    Support alarm review 

    Support alarm pause

     

    Linux OS

    Support operation with mouse and computer keyboard (option)

    Multi-display mode to meet different clinical requirement, including standard interface, big font, OxyCRG, trend graph, NIBP review and full leads ECG interface.

     

    NIBP self-test mode: including overpressure test, static pressure test and air leakage test.

    Generate ID automatically when register a new patient

    Support medical history search by patient ID, name and mobile number

    18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection

    Drug calculation and titration table 

    Support multi-language display

    Support online software upgrading by net/USB

     

    Performance Specifications

    Dimension and Weight

    Dimension: 380mm*360mm*160mm

    Weight: 4.5kg(excluding accessories)

     

    Power Supply

    Voltage: AC100~240V, 50/60HZ, Power≤60W

    Display

    15" color TFT LED  

    Resolution: 1024*768 pixels

    Battery (Pluggable)

    Type: Rechargeable lithium battery 14.8V/2200mAh

    Charge Cycle: ≥500 times

    Working time: 2 hours (optional larger battery for 4-5 hours)

    Recorder (Option)

    Method: Thermal printer

    Paper width: 50 mm (1.97 in)

    Printing speed: 12.5/ 25/ 50 mm/s

    Trace: Max. 3 tracks

    Recording way: Real-time Recording, Review Printing,  

    Periodic Recording, Alarm Recording

    Alarm

    Level: Low, medium and high

    Indication: Auditory and visual

    Alarm volume adjustable

    Alarm pause time: 1min, 2min

    Parameter alarm type: Latch/ Unlatch

     

    Input Device

    Touch screen (option)

    Knob (Standard)

    Keypad input (Standard)

    Mouse/ Keyboard input (option)

     

    System Output & Extensible Interface

    Ethernet Network: 1 standard RJ45 socket

    USB Port: 1

    Video Output:  1 HDMI port (option)

     

    Operating Environment

    Temperature:  5 ~ 40 ℃

    Humidity:  15% ~ 80% (non-condensing)

    Atmosphere pressure:  86 KPa ~ 110 KPa

    Transportation and Storage Environment

    Temperature: -20~50 ℃

    Humidity: 10%~90%  (non-condensing)

    Atmosphere pressure:  86 KPa ~ 110 KPa

     

    Safety

    IEC60601-1 Approved, CE marking according to MDD93/42/EEC

    With reference to RoHS Directive 2011/65/EU recasting

    Trend & Reviewing

    Trend: 2160 hours

    ARR events: 128 groups of ARR events and associated waveform

    NIBP measurement reviewing: 2000 groups

    Waveform review: 2 hours

    Alarm event: 2000 groups of parameter alarms events and associated parameter

    ECG

    Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V

    Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof

    Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto

    Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s

    ECG signal range:  ±12 mV (Gain × 0.25)

    Accuracy: ±1bpm/ ±1%, whichever is greater

    Resolution: 1 bpm

    Leakage Current < 10 μA

    Baseline Recovery:

    ≤ 3s after defibrillation (Monitor mode)

    ≤ 1s after defibrillation (Surgery mode)

    Bandwidth: Surgery 1 ~ 25 Hz

    Monitor 0.5 ~ 40 Hz

    Diagnostic 0.05 ~ 120 Hz

    Indication of Electrode Separation: Every electrode (exclusive of RL)

     

    Heart Rate

    Measure range:

    Adult: 0bpm; 10 ~ 300 bpm

    Neo/Ped: 0bpm; 10 ~ 350 bpm

    Resolution: 1 bpm

    Accuracy: ± 1% or ± 1bpm, whichever is greater

     

     

     

     

    ST Measurement

    Range: -2.0 ~ +2.0 mV

    Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater

    Other range: unspecified

    Resolution: 0.01mV

     

    Respiration

    Method: Impedance between RA-LL, RA-LA

    Gain: ×0.25, ×0.50, ×1, ×2, ×4

    Respiration Rate: 0bpm, 6 ~ 150 BrPM

    Sweep speed:  6.25 mm/s, 12.5 mm/s, 25mm/s

    Resolution: 1 BrPM

    Accuracy: ±2 bpm or ±2%, whichever is greater

    Apnea Alarm: 10 ~ 60 s

     

    NIBP

    Method: Oscillometric

    Measure mode: Manual, Auto, STAT

    Measure Interval in AUTO Mode 1~480 min

    STAT mode cycle time: Keep 5 minutes, at 5 seconds interval

    Measure and Alarm Range: 

    Adult: SYS:  30 ~ 270 mmHg

               DIA:  10 ~ 220 mmHg     

               MEAN:  20 ~ 235 mmHg

    Pediatric: SYS:  40 ~ 235 mmHg

    DIA:  10 ~ 220 mmHg     

    MEAN:  20 ~ 225 mmHg

    Neonate:   SYS:  30 ~ 135 mmHg

    DIA:  10 ~ 110 mmHg     

    MEAN:  20 ~ 125 mmHg

    Static pressure accuracy: ±3mmHg

    Resolution: 1mmHg

    Accuracy: Maximum Mean error ±5mmHg           

             Maximum Standard deviation ≤8mmHg

    Over pressure Protection: Dual protection via software & hardware

     

    Temperature

    Technique: Thermistor probe (2.25K)

    Channel:  Dual-channel, provide T1; T2; ΔT

    Measuring and Alarm Range:

    0.0 ℃~ 50 ℃ (32℉~ 122℉)

    Unit: Celsius (℃), Fahrenheit ( ℉)

    Resolution: 0.1℃or 1℉

    Accuracy: ±0.1℃(exclusive probe)

     

    SpO2 (Digital Technic)

    Measurement Range: 0 ~ 100 %

    Resolution: 1 %

    Response Modes: Low, Medium, High

    Accuracy:  ±2% (70% ~ 100%)

                ±3 % (35% ~ 69%)

    Unspecified (0 ~ 34%)

    Support Pitch tone and multi-level volume   

     

    Pulse Rate

    Measuring and Alarm Range: 25~250bpm

    Accuracy: ±1% or ±1 bpm, whichever is greater

    Resolution: 1bpm

     

    Nellcor-SpO2 (Option)

    Measurement Range: 0 ~ 100 %

    Resolution: 1 %

    Accuracy:   70% ~ 100%, ±2 % (adult)

                 70% ~ 100%, ±3 % (Neonate)

                 70% ~ 100%, ±2 % (Low Perfusion)

                 0% ~ 69%, unspecified

     

    Pulse Rate

    Measurement range: 20 ~ 300 bpm

    Resolution: 1bpm

    Accuracy:

    ±3 bpm (20 ~ 250 bpm)

    unspecified (251~300 bpm)

     

    Masimo SpO2 (Option)

    - Measurement Range: 0 ~ 100 %

    - Resolution: 1 %

    Accuracy:  

    70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions)

    70% ~ 100%, ±3 % (neonate, non-motion conditions)

    70% ~ 100%, ±3 % (motion conditions)

    0% ~ 69% unspecified

    Pulse Rate

    Measurement range: 25 ~ 240 bpm

    Resolution: 1bpm

    Accuracy: ±3 bpm (non-motion condition)

     

    EtCO2 (Mainstream/ Sidestream) (Option)

    Measure method:  Non-dispersive infrared (NDIR)

    Measure Range:  0 ~19.7% (0 ~ 150 mmHg)

                     0 ~ 20 kPa

    Resolution: 0.1 mmHg

    CO2 Accuracy:  

    0 ~ 40 mmHg, ±2 mmHg

    41 ~ 70 mmHg, ±5% of reading

    71 ~ 100 mmHg, ±8% of reading

    101~ 150 mmHg, ±10% of reading

    at 760 mmHg, ambient temperature of 25℃)

    Respiratory Rate:  Range: 3 ~150 BrPM

                    Accuracy: ±1 bpm

     

    EtCO2 (Micro-stream) (Option)

    Measure method:  Non-dispersive infrared (NDIR)

    Measure Range:  0 ~19.7% (0 ~150 mmHg)

                     0 ~ 20 kPa

    Sample Rate: 50 mL/min ±10mL/min

    Resolution:  0.1 mmHg (0 ~ 50 mmHg)

               0.25 mmHg (50 ~ 114 mmHg)

    CO2 Accuracy:  0 ~ 40 mmHg, ±2 mmHg

                    41 ~ 70 mmHg, ±5% of reading

                    71 ~ 100 mmHg, ±8% of reading

                   101~ 150 mmHg, ±10% of reading

            at 760 mmHg, ambient temperature of 35℃)

    Respiratory Rate: Range: 3 ~120 BrPM

                  Accuracy: ±1 bpm

     

    Depth of Anesthesia (CSI) (Option)

    EEG sensitivity:  ± 400μV

    Noise: < 2μVp-p, < 0.4μV RMS, 1-250 Hz



    Write down your demand below and we will contact you within 24h!
    ver_code
    Review

    ver_code
    1/3
    X