Core
Newly advanced A9 main board with Linux OS, calculating speed is 4 times faster than traditional products
Support storage of 2160 hours trend table and graph review, 2 hours waveform review, 2000 groups NIBP review and 2000 alarm events review
Body
15 inch high-brightness TFT LED
Support display 9~13 waveforms
Support 7 channel ECG waveform display simultaneously
Optional touch screen
Optional HDMI output
Printer
Built-in high-speed 50mm thermal printer
(Brand: SEIKO, Japan)
Central System
Optional Build-in wireless network module, supporting wired or wireless connection to the central monitoring station.
Optional Support HL7 (Health Level Seven)
Alarm
Three-level acousto-optic alarm
Sensor-off alarm
Paper out alarm
Support alarm review
Support alarm pause
Linux OS
Support operation with mouse and computer keyboard (option)
Multi-display mode to meet different clinical requirement, including standard interface, big font, OxyCRG, trend graph, NIBP review and full leads ECG interface.
NIBP self-test mode: including overpressure test, static pressure test and air leakage test.
Generate ID automatically when register a new patient
Support medical history search by patient ID, name and mobile number
18 types of Arrhythmia analysis and real-time S-T segment analysis and pacemaker detection
Drug calculation and titration table
Support multi-language display
Support online software upgrading by net/USB
Performance Specifications
Dimension and Weight
Dimension: 380mm*360mm*160mm
Weight: 4.5kg(excluding accessories)
Power Supply
Voltage: AC100~240V, 50/60HZ, Power≤60W
Display
15" color TFT LED
Resolution: 1024*768 pixels
Battery (Pluggable)
Type: Rechargeable lithium battery 14.8V/2200mAh
Charge Cycle: ≥500 times
Working time: 2 hours (optional larger battery for 4-5 hours)
Recorder (Option)
Method: Thermal printer
Paper width: 50 mm (1.97 in)
Printing speed: 12.5/ 25/ 50 mm/s
Trace: Max. 3 tracks
Recording way: Real-time Recording, Review Printing,
Periodic Recording, Alarm Recording
Alarm
Level: Low, medium and high
Indication: Auditory and visual
Alarm volume adjustable
Alarm pause time: 1min, 2min
Parameter alarm type: Latch/ Unlatch
Input Device
Touch screen (option)
Knob (Standard)
Keypad input (Standard)
Mouse/ Keyboard input (option)
System Output & Extensible Interface
Ethernet Network: 1 standard RJ45 socket
USB Port: 1
Video Output: 1 HDMI port (option)
Operating Environment
Temperature: 5 ~ 40 ℃
Humidity: 15% ~ 80% (non-condensing)
Atmosphere pressure: 86 KPa ~ 110 KPa
Transportation and Storage Environment
Temperature: -20~50 ℃
Humidity: 10%~90% (non-condensing)
Atmosphere pressure: 86 KPa ~ 110 KPa
Safety
IEC60601-1 Approved, CE marking according to MDD93/42/EEC
With reference to RoHS Directive 2011/65/EU recasting
Trend & Reviewing
Trend: 2160 hours
ARR events: 128 groups of ARR events and associated waveform
NIBP measurement reviewing: 2000 groups
Waveform review: 2 hours
Alarm event: 2000 groups of parameter alarms events and associated parameter
ECG
Lead mode: 3/5 Leads, I, II, III or I, II, III, AVR, AVL, AVF, V
Protection: Breakdown Voltage 4000VAC 50/60Hz; Defibrillator proof
Gain: 2.5mm/mV (×0.25), 5.0mm/mV (×0.5), 10mm/mV (×1), 20mm/mV (×2), 40 mm/ mV (×4), Auto
Sweep speed: 6.25mm/s, 12.5mm/s, 25mm/s, 50mm/s
ECG signal range: ±12 mV (Gain × 0.25)
Accuracy: ±1bpm/ ±1%, whichever is greater
Resolution: 1 bpm
Leakage Current < 10 μA
Baseline Recovery:
≤ 3s after defibrillation (Monitor mode)
≤ 1s after defibrillation (Surgery mode)
Bandwidth: Surgery 1 ~ 25 Hz
Monitor 0.5 ~ 40 Hz
Diagnostic 0.05 ~ 120 Hz
Indication of Electrode Separation: Every electrode (exclusive of RL)
Heart Rate
Measure range:
Adult: 0bpm; 10 ~ 300 bpm
Neo/Ped: 0bpm; 10 ~ 350 bpm
Resolution: 1 bpm
Accuracy: ± 1% or ± 1bpm, whichever is greater
ST Measurement
Range: -2.0 ~ +2.0 mV
Accuracy: -0.8mV~+0.8mV: ±0.02mV or ±10%, whichever is greater
Other range: unspecified
Resolution: 0.01mV
Respiration
Method: Impedance between RA-LL, RA-LA
Gain: ×0.25, ×0.50, ×1, ×2, ×4
Respiration Rate: 0bpm, 6 ~ 150 BrPM
Sweep speed: 6.25 mm/s, 12.5 mm/s, 25mm/s
Resolution: 1 BrPM
Accuracy: ±2 bpm or ±2%, whichever is greater
Apnea Alarm: 10 ~ 60 s
NIBP
Method: Oscillometric
Measure mode: Manual, Auto, STAT
Measure Interval in AUTO Mode 1~480 min
STAT mode cycle time: Keep 5 minutes, at 5 seconds interval
Measure and Alarm Range:
Adult: SYS: 30 ~ 270 mmHg
DIA: 10 ~ 220 mmHg
MEAN: 20 ~ 235 mmHg
Pediatric: SYS: 40 ~ 235 mmHg
DIA: 10 ~ 220 mmHg
MEAN: 20 ~ 225 mmHg
Neonate: SYS: 30 ~ 135 mmHg
DIA: 10 ~ 110 mmHg
MEAN: 20 ~ 125 mmHg
Static pressure accuracy: ±3mmHg
Resolution: 1mmHg
Accuracy: Maximum Mean error ±5mmHg
Maximum Standard deviation ≤8mmHg
Over pressure Protection: Dual protection via software & hardware
Temperature
Technique: Thermistor probe (2.25K)
Channel: Dual-channel, provide T1; T2; ΔT
Measuring and Alarm Range:
0.0 ℃~ 50 ℃ (32℉~ 122℉)
Unit: Celsius (℃), Fahrenheit ( ℉)
Resolution: 0.1℃or 1℉
Accuracy: ±0.1℃(exclusive probe)
SpO2 (Digital Technic)
Measurement Range: 0 ~ 100 %
Resolution: 1 %
Response Modes: Low, Medium, High
Accuracy: ±2% (70% ~ 100%)
±3 % (35% ~ 69%)
Unspecified (0 ~ 34%)
Support Pitch tone and multi-level volume
Pulse Rate
Measuring and Alarm Range: 25~250bpm
Accuracy: ±1% or ±1 bpm, whichever is greater
Resolution: 1bpm
Nellcor-SpO2 (Option)
Measurement Range: 0 ~ 100 %
Resolution: 1 %
Accuracy: 70% ~ 100%, ±2 % (adult)
70% ~ 100%, ±3 % (Neonate)
70% ~ 100%, ±2 % (Low Perfusion)
0% ~ 69%, unspecified
Pulse Rate
Measurement range: 20 ~ 300 bpm
Resolution: 1bpm
Accuracy:
±3 bpm (20 ~ 250 bpm)
unspecified (251~300 bpm)
Masimo SpO2 (Option)
- Measurement Range: 0 ~ 100 %
- Resolution: 1 %
Accuracy:
70% ~ 100%, ±2 % (adult/ pediatric, non-motion conditions)
70% ~ 100%, ±3 % (neonate, non-motion conditions)
70% ~ 100%, ±3 % (motion conditions)
0% ~ 69% unspecified
Pulse Rate
Measurement range: 25 ~ 240 bpm
Resolution: 1bpm
Accuracy: ±3 bpm (non-motion condition)
EtCO2 (Mainstream/ Sidestream) (Option)
Measure method: Non-dispersive infrared (NDIR)
Measure Range: 0 ~19.7% (0 ~ 150 mmHg)
0 ~ 20 kPa
Resolution: 0.1 mmHg
CO2 Accuracy:
0 ~ 40 mmHg, ±2 mmHg
41 ~ 70 mmHg, ±5% of reading
71 ~ 100 mmHg, ±8% of reading
101~ 150 mmHg, ±10% of reading
at 760 mmHg, ambient temperature of 25℃)
Respiratory Rate: Range: 3 ~150 BrPM
Accuracy: ±1 bpm
EtCO2 (Micro-stream) (Option)
Measure method: Non-dispersive infrared (NDIR)
Measure Range: 0 ~19.7% (0 ~150 mmHg)
0 ~ 20 kPa
Sample Rate: 50 mL/min ±10mL/min
Resolution: 0.1 mmHg (0 ~ 50 mmHg)
0.25 mmHg (50 ~ 114 mmHg)
CO2 Accuracy: 0 ~ 40 mmHg, ±2 mmHg
41 ~ 70 mmHg, ±5% of reading
71 ~ 100 mmHg, ±8% of reading
101~ 150 mmHg, ±10% of reading
at 760 mmHg, ambient temperature of 35℃)
Respiratory Rate: Range: 3 ~120 BrPM
Accuracy: ±1 bpm
Depth of Anesthesia (CSI) (Option)
EEG sensitivity: ± 400μV
Noise: < 2μVp-p, < 0.4μV RMS, 1-250 Hz