PCV and SIMV+PS modes ensure flexibility in ventilation, machine doable for both pediatric and adult operation
Optional monitoring functions, CO2 and anesthetic gas
Flowmeter provides at least 25% concentration of oxygen in O2 /N2O mixture
Breathing system easy to remove and clean, fully autoclavable, optional CO2 bypass
Enhances usability with auxiliary oxygen, suction, auxiliary power source, monitor arm, infusion pump etc.
CE marked, meets EU clinical requirement
|
Physical & Environment Specifications |
|
|
Dimensions(W×H×D) |
704mm × 1403mm × 750mm |
|
Weight |
100kg (basic unit) |
|
Storage Humidity: |
≤93% (noncondensing) |
|
Operation Humidity |
≤80% (noncondensing) |
|
Electrical Specifications |
|
|
Mains |
100~240V AC, 50/60Hz |
|
Battery |
24V DC, min 120 min |
|
Outlets |
3 on back, 1.5A individual |
|
Pneumatic Specifications |
|
|
Gas Supply |
Air, O2, N2O; 280kPa ~ 600kPa |
|
Flowmeters |
O2 Range: 0~1L/min; 1~10L/min Auxiliary O2 Range: 0 ~12 L/min |
|
Gas System |
Flowmeter backlight |
|
ACGO |
Optional auxiliary common gas outlet |
|
AGSS |
Optional anesthetic gas scavenging system |
|
Breathing System Specifications |
|
|
Operational Modes |
Closed, Semi-closed, Semi-open |
|
CO2 Canister |
2 Loose fill (3L) or 2 prepacks, |
|
APL Range |
Spontaneous breathing (SP) -70 cmH2O |
|
Material |
Autoclavable (except O2 cell and mechanical pressure gauge) |
|
CO2 bypas |
Optional |
|
Ventilator Specifications |
|
|
Driven Mode |
Pneumatically driven and electronically controlled |
|
Patient Types |
Adult, Pediatric |
|
Ventilation Modes |
IPPV, PCV, SIMV+PS, PS and Manual |
|
Setting Modes |
Touch key & navigator wheel knob |
|
Tidal Volume |
20~1500ml |
|
Ventilation Frequency |
4~100bpm |
|
I:E Ratio |
4:1~1:8 |
|
Pressure (inspired) Range |
5~70cmH2O |
|
Pressure (support) Range |
3~50cmH2O |
|
PEEP |
Off,3~30 cmH2O |
|
Tslope |
0~2s |
|
Inspiratory Plateau |
Off, 5%~60%Ti |
|
Trigger Sensitivity |
1~15L/min |
|
Compensation |
Compliance and leakage compensation, |
|
Ventilator Monitoring |
|
|
Display |
10.4” TFT color screen |
|
Pressure |
Ppeak, Pplat,Pmean,PEEP |
|
Volumes |
MV, Vt, Freq |









A1: The AEON8300A supports Volume-Controlled Ventilation (VCV), Pressure-Controlled Ventilation (PCV), SIMV, and Manual Ventilation, making it adaptable for various surgical scenarios.
A2: Yes, the AEON8300A features adjustable tidal volume settings and precise gas delivery, making it suitable for both pediatric and adult patients.
A3: Yes, this machine comes with built-in patient monitoring and can be integrated with external patient monitoring systems for enhanced safety and data tracking.
A4: The AEON8300A includes real-time monitoring of oxygen levels, tidal volume, and CO2 concentration, along with alarm systems for abnormal conditions.
A5: Regular maintenance includes filter replacements, gas line checks, software updates, and routine system calibration to ensure long-term performance.